Automatic Gas-Liquid Characterization Device

GL-1 Automatic Gas-Liquid Characterization Device

Information

Automatic Gas-Liquid Characterization Device

Innovative methods enable rapid gas-liquid equilibration, providing quick and user-friendly operation.

 

The original tube-in-tube microreactor achieves rapid gas-liquid equilibrium, facilitating swift measurements of gas-liquid properties using an online flow method. The system's software control feature automatically logs and stores data on temperature, pressure, liquid flow rate, and gas flow rate during measurements. Different working modes, such as quasi-steady and unsteady modes, are available with built-in flow calculation models in the software. The quasi-steady mode can determine gas solubility and diffusion coefficients in liquids, while the non-steady mode is ideal for fast diffusion coefficient measurements in high-viscosity systems like ionic liquids.

Features

  • Comprised of a high-precision gas flowmeter, a high-pressure syringe pump, a tube-in-tube microreactor, a back pressure valve (BPV), and system management software.

  • Using a dual continuous high-pressure syringe pump allows for a steady supply of liquid, especially for viscous solutions.

  • The membrane structure contains special materials that facilitate gas-liquid equilibration within 10 to 30 seconds, utilizing flow chemistry to swiftly determine gas-liquid properties.

  • The working mode is categorized into quasi-steady and unsteady modes. The integrated flow chemistry calculation model can rapidly determine gas solubility in liquid, diffusion coefficient, and reaction kinetics.


Specifications


    ModuleSpecifications
    PumpType: Dual-syringe high-pression pump
    Flow rate: 3 mL/min
    Accuracy: ±1 %
    Repeatability: RSD<0.5%
    Tube-in-tube
    Microreactor
    Material: inner tube: PTFE(AF2400); outer tube: 316L, C276, PFA
    Flow rate range: 0.001-3 mL/min
    Inner tube volume: 0.6 mL
    Middle layer volume: 5.2 mL
    Gas Mass Flow MeterFlow rate: 0.1- 5 sccm
    Accuracy: ±1 %
    Repeatability: RSD<0.5%
    Suitable gas: H2, N2, CO, CO2, NH3
    BackboneNo. of liquid pipeline: 2, 3
    No. of gas pipeline:1
    No. of microreactor: 1-4
    Working pressure: < 6 MPa
    Temperaure: -20-180 ℃
    Duration of solubility measurement:<5 min
    Duration of diffusion coefficient measurement:<5 min
    Duration of reaction kinetics measurement:<5 min
    Dimension (DxWxH, mm): 604x428x485