Ou shisheng Four-Channel Adsorption Reaction System: Parallel Screening and Multi-Stage Series Operation in One Unit
For research, development, and evaluation of adsorbents and catalytic materials, standard 4-channel reactors are primarily used for parallel screening. These reactors allow for simultaneous comparative experiments across multiple channels, speeding up the testing of samples and operating conditions. However, a common drawback of these setups is the inability to connect channels flexibly in series. This limitation makes it difficult to simulate real industrial processes that involve multi-stage beds and linked adsorption-reaction steps, such as hydrodesulfurization followed by hydrofining. Previously, conducting sequential process experiments required transferring materials between separate units. This inconvenient process often introduced disruptions to phase states, temperature, and concentration during transfer, compromising the accuracy of experimental data.
To overcome this industry challenge, Ou Shisheng has engineered a specialized 4-channel adsorption-reaction unit. While maintaining the full capacity for parallel multi-channel screening, the system utilizes reconfigurable valve manifolds to allow for flexible series connections between channels. A single instrument supports various operating modes: parallel screening, two-channel series connection, and full four-channel series connection. Featuring two independent sets of gas-liquid separation, back pressure regulation, and sampling modules, all four reactor positions can be connected to an on-line chromatography system. This platform effectively addresses two key experimental needs: high-throughput screening and continuous process simulation.

01 Parallel & Series Modes for Diverse Experimental Scenarios
The adsorption system consists of gas feed unit, liquid feed unit, reactor unit, automatic gas-liquid separation unit and automatic back pressure regulator unit. See the P&ID diagram below for detailed workflow.

Figure 1 | PID of Adsorption System
The system features four 316L reactors, each with a catalyst loading volume of 80 mL. Every reactor has its own electric furnace, allowing for independent temperature control.
The 4-channel parallel mode is a standard approach for high-throughput material screening. It allows for direct comparison of the same batch of adsorbent under different temperature and space velocity conditions. Alternatively, with consistent overall process parameters, four different adsorbents can be tested at the same time. This condenses screening tests that would normally require four separate runs into a single operation.
Using valve-manifold flow switching, the system can also be reconfigured into two series modes: a two-pair series and a full four-channel series. In the full four-channel series mode, the outlet stream from one reactor flows directly into the next. This eliminates intermediate sampling and material transfer, preventing disruptions to the phase and parameters that can occur when streams are cooled and reheated. Continuous material flow between beds is maintained, with each stage holding stable, independently set temperatures. Coupled adsorption-reaction sequences are contained within a single unit, without material transfer between different pieces of equipment, accurately reflecting real continuous process operating conditions.
The two-pair series configuration divides the four reactors into two completely independent series setups. Both setups can run experiments simultaneously, or one can be used on its own. Process parameters for the two flow paths can be adjusted independently without affecting each other, enabling multiple parallel comparative tests. Unlike traditional 4-channel systems that are limited to parallel single-column adsorption tests, Ou Shisheng’s system enhances experimental capabilities. It supports standard single-stage adsorption performance evaluation and also allows for the simulation of industrial multi-stage bed conditions. This single platform is ideal for rapid adsorbent screening and iterative process optimization, accommodating a wider range of process designs and reducing the capital costs associated with purchasing or modifying separate equipment for different configurations.
02 Series Operation Assurance: Preventing Material Balance Distortion by Suppressing Cold Spots in Piping
For gas-phase experiments conducted in series, cold spots in the pipelines present the main engineering challenge. When vaporized feed leaves the upstream reactor, if the temperature of the inter-column pipeline drops below the stream's dew point, heavier components can condense on the pipe walls and fail to reach the downstream reactor. This directly distorts the material balance between the reactors, rendering the experimental data unreliable.
Ou Shisheng’s 4-channel adsorption-reaction unit incorporates a targeted thermal management design. It is equipped with two vaporization chambers and two gas preheaters. The segmented, heat-traced lines extending from the vaporization chambers to the reactor inlets are individually temperature-controlled. Furthermore, all connecting lines between reactors and those leading to the inline chromatograph are fitted with heat tracing and thermal insulation. Pipeline temperatures can be set independently for different feeds, meaning that test substances such as n-heptane, toluene, cresol, and dimethylnaphthalene do not have to share a single pipeline temperature setting.
The process principle for liquid-phase adsorption differs from that of the gas phase. Multiple liquid feeds are metered and blended according to a specific formulation before entering the reactor directly. In this case, the piping only needs to maintain a stable feed state and deliver the fluid without pulses, rather than sustaining a high-temperature gas zone. This is why the gas and liquid circuits are configured independently: the gas circuit focuses on precise temperature control and preventing condensation, while the liquid circuit prioritizes accurate proportioning and low-pulse delivery. The liquid circuit comprises four high-pressure plunger pumps. In addition to the two main feeds, diluents or spare pumps can be integrated for flexible distribution as required by the tests.
The hardware of the feed delivery unit further enhances experimental accuracy. The liquid circuit utilizes four high-pressure plunger pumps with a flow rate range of 0.001–9.999 mL/min and a control precision of 0.001 mL/min. The gas circuit is equipped with two bypass-type mass flow controllers, compatible with nitrogen and hydrogen, covering a range of 5–1000 mL/min, to ensure stable gas and liquid feeding.
03 Split Outlet Design for Differential Pressure Control in Series Operation
Traditional 4-channel systems use a single outlet processing unit for all channels. In series experiments, this arrangement forces the upstream and downstream beds to operate under identical pressure conditions. This limitation prevents the simulation of real industrial processes where pressure drops or varies along the process flow.
This system, however, divides the outlet processing assembly into two independent sections. Each reactor bank is equipped with its own gas-liquid separation and back pressure module, allowing for separate pressure settings. This enables upstream and downstream beds to operate under different pressure conditions during series workflows.
Each gas-liquid separator has a volume of 100 mL and is constructed from 316L stainless steel, designed for a maximum pressure of 10 MPa. The liquid level can be adjusted from 2% to 95% of the full scale with a control accuracy of ±1%. Continuous liquid discharge is achieved using an automatic high-pressure constant-flow control valve. Pressure regulation is managed by diaphragm-type automatic back pressure valves, which offer a control accuracy of ±0.1 MPa, pressure repeatability (RSD) of less than 0.5%, and a valve response time of less than 3 seconds.
For safety, the unit includes multi-layered protection: power cutoff in case of overtemperature, pump shutdown if overpressure occurs, pressure relief through safety valves, and an emergency stop button. For systems using hydrogen, instruments and junction boxes are selected and installed in compliance with hazardous area classifications for hydrogen environments.
04 Traceability for All Experimental Results
To facilitate the review and traceability of experimental findings, the system incorporates two analytical workflows: off-line sampling and inline gas chromatography.
A 16-position automatic sampling module is installed downstream of each gas-liquid separation and back pressure unit, with each sample volume ranging from 0.5–1.5 mL. Sampling sequences are synchronized with process conditions, allowing for continuous sampling according to preset time schedules. The module features optical recognition of vial positions and automatic line rinsing to minimize cross-contamination and prevent sampling errors caused by air bubbles. Collected liquid samples can be stored for subsequent off-line verification. Four sampling ports are strategically placed at different process points on the inline GC system, capturing compositional changes before and after reaction, as well as before and after separation.
Dedicated control software synchronously acquires pressure, flow rate, temperature, and other parameters along a unified timeline. All data are archived batch by batch, with operating parameters precisely matched to chromatographic results. Real-time equipment status can be monitored remotely via mobile phones or tablets. Every dataset is fully retrievable and verifiable, offering complete and reliable data support for mechanism investigation and iterative process optimization.